Overview
This article describes water-sample calibration for ZIRKON® ONE DIS on NEON® ONE / KRYPTON® ONE.
Procedure
- Open Sensor Details > Cal.
- Select Water Sample Calibration.
- Take sample near measuring point.
- Measure reference value (DPD), ideally 3 repeats and average.
- Enter value and confirm calibration.
Technical Note
NEON® ONE uses a slope correction factor (default around 1.0), not legacy mV/ppm display logic.
Important
- Calibrate only with stable flow.
- Do not calibrate during ASR cleaning or immediately after ASR.
- Use clean sensor surfaces before calibration.