What to do if you receive a mA out of range message on the analyzer.
Typical Result:
- Most Common- Measurement exceeds selected measuring range
- Common- mA output settings- improper scaling
mA Out of Range
Cause: The current measured value corresponds to an output outside the 0(4)-20 mA range
Suggested remedies: Check settings
For example: measured value 5.6ppm, mA set to 4mA=0ppm; 20mA = 5ppm
Please pay attention to the selected measuring range and mA output settings compared to the current measured value and expected min./max. concentration of the process.
Note - The Kuntze disinfectant measuring range is selected in the instrument menu. Available measuring ranges are . . .
0 - 1000 µg/L
0 - 5.00 mg/L
0 - 10.00 mg/L
0 - 20.00 mg/L
To change the selected measuring range go to . . .
Main Menu > In/out > Analog > DIS 1 > selected measuring range.