What to do when you receive a load impedance error on the analyzer.
Typical Result:
- Most Common- The mA output is not connected to an aux. device. i.e. PLC, dosing pump, data logger, etc.
- Common- The mA output wiring is compromised
- Least Common- Krypton® Multi hardware has been compromised or damaged
Load impedance error
Cause: A mA output has been activated, but is either unconnected or the connection is broken
Suggested remedies: Check connection or change the assignment of the mA output
Main Menu > IN/OUT > mA OUTPUT > (Select 1 of the 5 outputs that is in alarm)
Note: The instrument monitors the Ohmic load of the mA outputs and can detect if a connection is broken. If an mA output is not connected at all because you do not want to use it, change the mA output assignment to "not assigned"