Kuntze analyzers let you store configuration on an SD card for backup or to copy settings to another unit of the same type. This applies to Krypton® Multi, Krypton® DIS, and Krypton® DIS (Total) (same Memory workflow as DIS; instrument type must match when loading).
Overview
| Topic | Krypton® Multi | Krypton® DIS / DIS (Total) |
|---|---|---|
| Typical SD access | SD in lid assembly (snap-on lid) | SD on rear (four corner screws, turn unit over) |
| Save path (symbols) | SAVE SETTINGS > OK | SAVE > SETTINGS > OK |
| Load path (symbols) | LOAD SETTINGS > OK | LOAD > SETTINGS > OK |
You can restore settings after a replacement instrument, a firmware update, or after enabling add-ons. With a suitable PC tool you can edit settings files offline and load them via SD (same type restrictions apply).
Note: Settings may only be duplicated or loaded onto instruments of the same type (Multi → Multi; DIS → DIS; DIS Total → DIS Total).
Before you begin
- Industry-standard SD card, maximum 2 GB, with free space.
- Card installed in the analyzer before save.
- For physical access: Multi — open lid; DIS / DIS (Total) — rear panel per maintenance articles.
Krypton® Multi — save and load
Save settings
- Confirm the SD card is inserted and has free capacity.
- Main menu > IN / OUT > MEMORY — select SAVE SETTINGS > OK.
- The display shows the new filename (parameter and date). Confirm with OK.
Load settings
- In MEMORY, select LOAD SETTINGS > OK.
- Confirm the filename shown (parameter and date) with OK.
- The instrument restarts; time is shown during startup.
Krypton® DIS / DIS (Total) — save and load
Save settings
- Confirm the SD card is inserted and has free capacity.
- Main menu > IN / OUT > MEMORY — select SAVE > SETTINGS > OK.
- The display shows the new filename (parameter and date). Confirm with OK.
Load settings
- In MEMORY, select LOAD > SETTINGS > OK.
- Confirm the filename shown (parameter and date) with OK.
- The instrument restarts; time is shown during startup.
Note: Load only files intended for the same parameter / instrument family.
Important notes
- Do not mix settings files between Multi and DIS platforms.
- After major changes (firmware, factory reset), reload a saved file only if it matches the new software and options enabled on the device.
- Consider saving settings before firmware updates or service.